health

The slip of paper that made an eye-pressure gauge readable

4 sources 3 primary sources November 25, 2026

Loading reads and saves…
Text

Behind the padded lid of a German-made eye instrument from about 1930, there is a slip of paper. The Australian College of Optometry’s museum describes a cream-colored scale with tick marks, a set of small weights with one missing, and, on the hidden sheet, a separate conversion curve for each load. A physician who had watched the pointer settle still had to look inside the lid to finish the examination.[1]

What the physician wanted was the pressure inside the eye a moment before the instrument touched it. By the time the pointer gave a reading, that pressure had already changed.[2]

A Schiøtz eye tonometer and its small metal fittings in an open, velvet-lined carrying case.
The German-made set, dated to about 1930. The recesses kept the weights with the instrument that had been calibrated to use them. Photograph from the Australian College of Optometry, catalogue 2891.[1]

What moved the pointer

Hjalmar Schiøtz demonstrated his tonometer to the Norwegian Medical Society in 1905. The patient lay back, face upward. A curved footplate rested on the numbed cornea, the clear front surface of the eye. Through the middle of the footplate ran a weighted plunger. Gravity drew it down, the cornea yielded beneath it, and a lever magnified that small movement into a sweep of the pointer. With the same weight, a softer eye let the plunger sink farther. A larger reading on the scale therefore meant a lower estimated pressure.[1][2][3]

A later manufacturer's manual gives a sense of the distances involved. One division of its scale represents just 0.05 millimeters of plunger travel. Imagine finding a penciled “5” in an old examination record. In the manual’s 1955 conversion table, five divisions under a 5.5-gram load correspond to 17.3 millimeters of mercury. Under a 7.5-gram load, the very same five divisions give 25.8. An eye that yielded only that far under the heavier load had offered more resistance. Without the weight, the penciled number has lost half its meaning.[3]

What the eye did under the weight

Press a small hollow into the surface of a fluid-filled eye and the fluid beneath it has to go somewhere. The wall of the eye stretches to accommodate the displacement. How much pressure rises depends partly on how readily that wall stretches. Two eyes could start at the same pressure and resist the instrument differently. A stiff eye could appear to have started at a higher pressure than it did.[2][3]

In 1947, ophthalmologist Jonas Friedenwald set out a way of relating the pressure under load to the pressure at rest. One calibration experiment began with an eye connected to a pressure gauge by a fluid line. The experimenter could record the starting pressure, close the line, then lower the instrument and read the scale. In a second arrangement, the line stayed open, allowing displaced fluid to escape and its volume to be measured. Closing or opening a tap separated two quantities that a doctor examining a patient could never see separately.[2]

In 1967, U. Majzoub at the University of London tried a transparent tonometer. Looking through it at a mounted human cornea, Majzoub saw a narrow gap between part of the footplate and the tissue. As the experimenter raised the fluid reservoir to increase pressure, the gap shrank. Lowering the reservoir let it widen again. Even the surface against which the plunger's travel was measured could shift. Majzoub proposed a lighter footplate whose curve more closely followed the cornea, so the act of measurement would displace less fluid.[2]

Another surviving Schiøtz set belonged to Montreal physician A. Isserman. The Canadian museum Ingenium records that the instrument itself is complete, but its table of values is gone.[4] The needle can still tell a visitor how far the metal moved. The missing paper held the other part of the examination, the experimental work that allowed a doctor to look at a tiny dent and estimate the pressure in an undisturbed eye.[2][4]

Sources

  1. Australian College of Optometry, “Schiotz tonometer,” catalogue 2891. Object description, photograph and provenance.
  2. U. Majzoub, “Volume Changes in Indentation Tonometry,” British Journal of Ophthalmology, 1967. Calibration experiments and observations with a transparent tonometer.
  3. Rudolf Riester, directions for the Schiøtz tonometer, including the 1955 conversion table and scale specifications.
  4. Ingenium – Canada’s Museums of Science and Innovation, tonometer used by Dr. A. Isserman, artifact 2002.0462.001.
Previous After a baby’s cataract comes out, can the implant wait? Next What could a week less breathing support mean at age eleven?

Recommended In health

Matched by subject and format