A blood-pressure cuff closes around the ankle. A pen-shaped ultrasound probe rests against the skin, turning arterial blood flow into a sound the examiner can hear. Cambridge University Hospitals tells patients to expect swooshing noises during this examination, which compares pressures in the upper arms and ankles.[1]
The result is the ankle–brachial index, usually shortened to ABI; British services often call it ABPI. A low ratio can reveal impaired circulation to the leg. But the scale has a trap: an unusually high result can also be abnormal. To understand why, follow what happens between the cuff tightening and the number appearing on the report.[1][4]
The sound supplies the pressure reading
The cuff compresses the artery until its flow signal disappears. As cuff pressure falls, blood passes again. The pressure at which the Doppler signal returns is the systolic reading.[3]
That distinction matters because hearing blood flow alone does not establish that circulation is adequate. The examination measures the pressure needed to interrupt and then permit that flow. Cambridge's patient leaflet describes measurements at the arms and ankles, without needles or radiation.[1]
Stanford Medicine specifies resting flat for 10 minutes, a correctly sized cuff, and measurements from both arms and two arteries at each ankle.[3]
What the division preserves
Each leg gets its own ABI: the higher pressure from that ankle's two arteries divided by the higher arm pressure.[3]
For illustration, 90 mmHg at the ankle divided by 120 mmHg at the arm gives 0.75. This compares pressures, not the percentage of artery remaining open.[3] Cambridge's explanation connects a lower ankle pressure with narrowing or blockage in the leg arteries.[1]
The comparison preserves the side of the problem: a low value on one leg remains visible even if the other leg has a normal result.
The 2024 ACC/AHA multisociety guideline separates resting results into four categories: 0.90 or below is abnormal; 0.91–0.99 is borderline; 1.00–1.40 is normal; and above 1.40 is noncompressible.[5] The final category is the reason “higher” cannot simply mean “better.”
When the artery resists the cuff
The measurement assumes the artery will yield to outside pressure. A stiff, calcified wall can resist the squeeze, producing an artificially high reading.[3][4]
The American Heart Association's 2012 statement connected ABI values above 1.40 with stiff arterial walls and possible calcification in the middle layer of the wall. It also emphasized standardized measurement and interpretation. The difficulty was already recognized as part of the test's meaning, rather than an unusual exception to ignore.[4]
There are consequently two different questions inside an abnormal report. A low ratio raises concern about pressure reaching the ankle. A noncompressible result raises concern about whether cuff pressure is faithfully measuring pressure inside the vessel. The arithmetic still produces a number when the second assumption fails; arithmetic cannot repair the measurement.
For suspected peripheral artery disease with ABI above 1.40, the 2024 guideline recommends toe pressure or toe–brachial index testing with waveforms. Toe arteries are rarely noncompressible, offering another measurement site when the ankle result is unreliable.[5]
A resting leg may leave the question open
Someone whose leg hurts while walking brings a question about activity to a test performed at rest. The National Heart, Lung, and Blood Institute describes exercise ABI testing as a way to examine symptoms and pressure after treadmill walking. It also explains why the history matters: joint, vein and nerve problems can produce leg pain that resembles arterial disease.[6]
The 2024 guideline recommends treadmill ABI testing for suspected chronic symptomatic disease with exertional, nonjoint leg symptoms despite a normal or borderline resting ABI.[5] Resting measurements leave this activity-dependent question open.
Other investigations answer different questions. NHLBI describes ultrasound imaging that can locate reduced flow or blockages, and segmental pressure measurements that examine different parts of the leg.[6] The ABI begins an assessment; the next test depends on what remains unexplained.
Finding disease and proving a screening benefit
A useful diagnostic test does not automatically justify testing everybody. In 2018, the US Preventive Services Task Force found insufficient evidence to determine the balance of benefits and harms of ABI screening in asymptomatic adults without known peripheral artery disease, cardiovascular disease or severe chronic kidney disease. That judgment concerned whether screening improves outcomes in that population.[7]
The 2024 multisociety guideline considers screening reasonable for people at increased risk, while recommending against it in those without increased risk and without suggestive findings.[5]
A pressure ratio becomes useful when its measurement limits and the reason for ordering it remain attached to the number.
Sources
- Cambridge University Hospitals, “Ankle brachial pressure indices test,” patient information, approved November 13, 2025 — examination setting, sounds and pressure comparison.
- Helios Klinikum Emil von Behring, vascular center — source of the ankle examination photograph credited to Dirk Pagels.
- Stanford Medicine 25, “Measuring and Understanding the Ankle Brachial Index (ABI)” — Doppler technique, preparation, calculation and noncompressible vessels.
- American Heart Association, “Top Things to Know: Measurement and Interpretation of the Ankle-Brachial Index,” November 16, 2012 — official summary of the scientific statement.
- Gornik and colleagues, 2024 ACC/AHA multisociety guideline for lower-extremity peripheral artery disease, section 3.1 — primary guideline PDF hosted by the International Society of Cardiovascular Pharmacotherapy.
- National Heart, Lung, and Blood Institute, “Peripheral Artery Disease: Diagnosis” — exercise testing, alternative causes of pain and further vascular investigations.
- US Preventive Services Task Force, final recommendation on ABI screening and cardiovascular risk assessment, July 10, 2018 — population and limits of the screening evidence.